Abstract
In this paper a new thermal imaging system based on a silicon CCD-camera is presented. Aim of the project is to monitor the surface of parts during a thermo-chemical heat treatment process as nitriding. The measurement system is based on the principles of ratio pyrometry known from non-contact temperature measurements. Two images of the same scene taken in slightly different spectral bands are used to determine the temperature of the surface. A black body radiator is used as reference for calibration of the camera-system. To enhance accuracy of the system correction of quantum efficiency and non linearity of the camera are corrected
Translated title of the contribution | Messsystem zur Prozesskontrolle bei Wärmebehandlungsprozessen auf Basis einer CCD-Kamera |
---|---|
Original language | German |
Title of host publication | 1. Forschungsforum der österreichischen Fachhochschulen |
Subtitle of host publication | Tagungsband |
Number of pages | 2 |
Publication status | Published - 2007 |
Event | Erstes Forschungsforum der österreichischen Fachhochschulen - Fachhochschule Salzburg, Campus Urstein, Austria Duration: 11 Apr 2007 → 12 Apr 2007 |
Conference
Conference | Erstes Forschungsforum der österreichischen Fachhochschulen |
---|---|
Country/Territory | Austria |
City | Fachhochschule Salzburg, Campus Urstein |
Period | 11.04.2007 → 12.04.2007 |