Abstract
We present multimodal noncontact photoacoustic (PA) and optical coherence tomography (OCT) imaging. PA signals are acquired remotely on the surface of a specimen with a Mach-Zehnder interferometer. The interferometer is realized in a fiber-optic network using a fiber laser at 1550 nm as the source. In the same fiber-optic network, a spectral-domain OCT system is implemented. The OCT system utilizes a supercontinuum light source at 1310 nm and a spectrometer with an InGaAs line array detector. Light from the fiber laser and the OCT source is multiplexed into one fiber using a wavelength-division multiplexer; the same objective is used for both imaging modalities. Reflected light is spectrally demultiplexed and guided to the respective imaging systems. We demonstrate two-dimensional and three-dimensional imaging on a tissue-mimicking sample and a chicken skin phantom. The same fiber network and same optical components are used for PA and OCT imaging, and the obtained images are intrinsically coregistered.
| Originalsprache | Englisch |
|---|---|
| Aufsatznummer | 046013 |
| Fachzeitschrift | Journal of Biomedical Optics |
| Jahrgang | 20 |
| Ausgabenummer | 4 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - 1 Apr. 2015 |
Fingerprint
Untersuchen Sie die Forschungsthemen von „Multimodal noncontact photoacoustic and optical coherence tomography imaging using wavelength-division multiplexing“. Zusammen bilden sie einen einzigartigen Fingerprint.Zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver