Membrane Focusing-A New Injection Method for Capillary Gas Chromatography

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

3 Zitate (Scopus)

Abstract

The principles of a new sample focusing technique in capillary gas chromatography are described. A solid-sorbent trapped sample is thermally released from the sample tube, transfered to a membrane chamber by a carrier gas and there retained by a gas separation membrane which is highly permeable to the carrier gas. After focusing in the membrane chamber the sample is injected on the column. This technique allows focusing and injection of highly volatile compounds in capillary GC without any coolant.

OriginalspracheEnglisch
Seiten (von - bis)1375-1387
Seitenumfang13
FachzeitschriftAnalytical Letters
Jahrgang22
Ausgabenummer5
DOIs
PublikationsstatusVeröffentlicht - Apr 1989
Extern publiziertJa