Matrix segregation as the major cause for sample inhomogeneity in MALDI dried droplet spots

Stefan J. Gabriel, Clemens Schwarzinger, Bettina Schwarzinger, Ulrich Panne, Steffen M. Weidner

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

31 Zitate (Scopus)

Abstract

The segregation in dried droplet MALDI sample spots was analyzed with regard to the matrix-to-sample ratio using optical microscopy, MALDI imaging mass spectrometry (MALDI MSI) and IR imaging spectroscopy. In this context, different polymer/matrix/solvent systems usually applied in the analysis of synthetic polymers were investigated. The use of typical matrix concentrations (10 mg mL-1) in almost every case resulted in ring patterns, whereas higher concentrated matrix solutions always led to homogeneous sample spot layers. The data revealed that segregation is predominantly caused by matrix transport in the drying droplet, whereas polymer segregation seems to be only secondary.

OriginalspracheEnglisch
Seiten (von - bis)1356-1363
Seitenumfang8
FachzeitschriftJournal of the American Society for Mass Spectrometry
Jahrgang25
Ausgabenummer8
DOIs
PublikationsstatusVeröffentlicht - Aug. 2014