Abstract
We propose an online method to estimate the time-dependent time-delay of continuous glucose measurement devices. The problem is characterized by one delayed signal which is sampled continuously, and a reference signal which is only available for measurement at (few) irregular time samples. The method makes use of a forgetting function to be able to quickly adapt to changes in the time-delay. A simulation example and the application to blood glucose concentration data obtained from type 1 diabetic subjects, where it is of interest to estimate the time-delay of the continuous glucose measurement device, show the usefulness and effectiveness of the proposed method.
Originalsprache | Englisch |
---|---|
Aufsatznummer | 7040486 |
Seiten (von - bis) | 6981-6986 |
Seitenumfang | 6 |
Fachzeitschrift | Proceedings of the IEEE Conference on Decision and Control |
Jahrgang | 2015-February |
Ausgabenummer | February |
DOIs | |
Publikationsstatus | Veröffentlicht - 2014 |
Extern publiziert | Ja |
Veranstaltung | 2014 53rd IEEE Annual Conference on Decision and Control, CDC 2014 - Los Angeles, USA/Vereinigte Staaten Dauer: 15 Dez. 2014 → 17 Dez. 2014 |