LMI-Based Approaches for the Calibration of Continuous Glucose Measurement Sensors

Harald Kirchsteiger, Luca Zaccarian, Eric Renard, Luigi Del Re

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

13 Zitate (Scopus)

Abstract

The problem of online calibration and recalibration of continuous glucose monitoring (CGM) devices is considered. Two different parametric relations between interstitial and blood glucose are investigated and constructive algorithms to adaptively estimate the parameters within those relations are proposed. One characteristic is the explicit consideration of measurement uncertainty of the device used to collect the calibration measurements. Another feature is the automatic detection of fingerstick measurements that are not suitable to be used for calibration. Since the methods rely on the solution of linear matrix inequalities resulting in convex optimization problems, the algorithms are of moderate computational complexity and could be implemented on a CGM device. The methods were assessed on clinical data from 17 diabetic patients and the improvements with respect to the current state of the art is shown.

OriginalspracheEnglisch
Aufsatznummer6867284
Seiten (von - bis)1697-1706
Seitenumfang10
FachzeitschriftIEEE Journal of Biomedical and Health Informatics
Jahrgang19
Ausgabenummer5
DOIs
PublikationsstatusVeröffentlicht - 1 Sep. 2015
Extern publiziertJa

Fingerprint

Untersuchen Sie die Forschungsthemen von „LMI-Based Approaches for the Calibration of Continuous Glucose Measurement Sensors“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren